| Capacitance diaphragm gauge | About 10−5 to 103 Torr, depending on the full-scale model | Measures diaphragm deflection caused by pressure; the reading is largely independent of gas type. | High accuracy and repeatability; useful for process control and calibrations; suitable for many corrosive or condensable gases when the wetted materials are compatible. | Each sensor has a finite range; low-range versions can be costly and may require temperature control for best stability. | Applications where accurate, gas-independent pressure measurement matters. |
| Pirani gauge | Approximately 10−4 to 103 Torr, depending on design | Infers pressure from heat loss from a heated element to the surrounding gas. | Broad rough- and medium-vacuum coverage; robust and commonly used for pump-down monitoring. | Response depends on gas composition and temperature; contamination can affect the sensing element and calibration. | General-purpose vacuum systems, chamber monitoring, and foreline measurement. |
| Thermocouple gauge | Typically about 10−3 to 1 Torr | Uses the temperature of a heated filament, sensed by a thermocouple, as an indication of gas heat conduction. | Simple, relatively economical, and useful in the rough-to-medium vacuum region. | Gas-dependent and generally less accurate than a capacitance diaphragm gauge; not suitable for high-vacuum measurement. | Basic vacuum indication and pump-down checks where moderate accuracy is sufficient. |
| Cold-cathode ionization gauge | Approximately 10−2 to 10−9 Torr; the usable range depends on the design and conditions | Uses a discharge in a magnetic field to ionize gas; ion current is related to pressure. | No hot filament; can measure high vacuum and is often mechanically robust. | May not start reliably at higher pressures; readings depend on gas and surface condition; discharge can be affected by contamination. | High-vacuum systems that need a filament-free ionization gauge. |
| Hot-cathode ionization gauge | Approximately 10−2 to 10−10 Torr; the lower limit varies with instrument and setup | Electrons emitted by a heated filament ionize gas molecules; the resulting ion current indicates pressure. | Sensitive and widely used for high- and ultrahigh-vacuum measurement; offers a broad low-pressure range. | Filament can be damaged by excessive pressure or reactive gases; readings depend on gas species, and very low readings can be limited by X-ray effects. | Clean high-vacuum and ultrahigh-vacuum systems requiring sensitive pressure monitoring. |
| Spinning-rotor gauge | Typically about 10−2 to 10−7 Torr, depending on the instrument | Determines pressure from the rate at which gas molecules slow a magnetically suspended rotating sphere. | Good repeatability and useful as a transfer standard; no hot filament or ionization discharge. | Typically slower to respond and sensitive to gas species, vibration, and sphere condition; less suitable for fast-changing pressure. | Calibration work and stable vacuum measurements where repeatability is important. |